PR 5500 Instrument Manual
Commissioning
Sartorius EN-165
5.11.8 Calibrating with Weight [Max at]
To calibrate the scale with weights:
t
t
Press the [by load] soft key.
t
t
Place the calibration weight on the scale.
t
t
Enter the weight of the calibration weight.
t
t
Confirm the entries.
The weight unit for the calibration weight (double
click to change) may differ from the unit in the
device; conversion is automatic.
Afterward, the following message is displayed:
“setting SPAN by load...”
The weight value of the calibration weight, weight
unit, and measuring signal in mV/V corresponding to
this value are displayed in the [calibrated at] line.