Commissioning
PR 5500 Instrument Manual
EN-270 Sartorius
5.19.6.5 EtherNet IP Settings for Rockwell Workstation
Requirements
-
The PR 1721/67 EtherNet-IP fieldbus card is installed; see Chapter 4.4.12.
-
The parameters have been selected and saved; see Chapter 5.19.6.1.
Procedure
t
t
Establish communication with the workstation in the ‚RSLinx‘ program
(initial ser.DF1, recommended ETH-LAN).
t
t
Register the file ‚sag_5500_ETH.EDS‘ in the ‚RSNetWorx for EtherNetIP‘ program.
t
t
Add a database entry in the hardware window under
[Vendor] - [Sartorius Mechatronics T&H GmbH] -
[Communications Adapter]:
Drag ‚MAXXIS Device‘ to ‚Graph‘ and apply the ETHIP address specified on the instrument.
t
t
Start the ‚RSLogix5000‘ program.
t
t
Upload from the workstation to the project newly created with the <RSLogix5000> program.
t
t
In the <RSLogix5000> program, set the access to ‚Online‘ in
[Controller Tags] as [DINT per WP 2 lines
with 4 bytes each].
t
t
Additional settings are defined in the application in question, such as phase (see corresponding manual).
Card Test
The fieldbus card has 3 LEDs (MS, Link, NS). Communication is functioning correctly when the LEDs are green
(see Chapter 4.4.12).
All inputs and outputs are displayed under [System maintenance] - [Test hardware] - [I/O card test] -
[Option-FB PR 1721/67].
Note
Further details can be found in the supplementary application manual ‚How to...‘ (available
upon request from [email protected]).
‚RSLinx‘, ‚RSNetWorx‘, and ‚RSLogix5000‘ are products of Rockwell Automation®.