Commissioning
PR 5500 Instrument Manual
EN-158 Sartorius
5.11.4.2 Modifying a Calibration
Example
Resetting the Dead Load
t
t
Select and confirm [System setup]-[Weighing
point]-[Weighing point A]-[Calib].
t
t
Press the [Modify] soft key.
t
t
Select the menu item [Dead load at].
t
t
Either press the [by mV/V] soft key to enter the
value again or clear the scale/hopper and press
the [by load] soft key to reset the dead load.
t
t
Press the [Save] soft key to save the calibration.
t
y
The reference value is calculated for the
electronics test.