Extended Functions
PR 5500 Instrument Manual
EN-298 Sartorius
t
y
An input window appears.
t
t
Enter a comment using the keyboard.
t
t
Press the [Continue] soft key.
t
y
The progress windows for the individual items
appear and then disappear in sequence.
t
y
As soon as the process is complete, an info
window appears.
t
t
Press the [OK] soft key to return to the [System
maintenance] menu.
6.1.5 Importing Data from External Media
This function is required to import the exported data (or e.g. database tables generated by the PC) on device.
Data to be imported must be in XML format. The format must be the same as that of exporting. The data can
be created manually (using e.g. Windows Notepad) or via software export with known format.
Examples
-
Creating operating parameters to make the parameters for all devices available in the system.
-
Providing data for user management to ensure the same access rights for all devices.
-
Providing a new tare table every day for the BASIC application.
Important
During the importing process, only data present in the data set being imported will be
overwritten.
The following applies:
-
Importing the setup data will only replace the settings present in the data set. For
example, if the data set only contains operating parameters, the other settings will
remain unchanged.
-
Importing the calibration data will only replace the calibration values present in the
data set.
-
Importing the user management data will only replace the settings present in the data
set. For example, if the data set only contains new users, the existing users will remain
unchanged.
-
Importing the application data will only replace the settings present in the data set. For
example, if the data set only contains print parameters, the other settings will remain
unchanged.
-
Importing the database tables will only replace the tables present in the data set. If the
data set contains the tare table only, the text table will remain unchanged in the device;
the tare table will be overwritten completely however.