5.14.3 Calibration Sequence
During calibration, no data is changed in the digital load cells. The calibration data and parameters are saved in
the instrument. The unique serial numbers of the connected load cells are monitored. The number of platforms
greatly influences how calibration is carried out.
For the adjustment is the following order specified:
-
Search for load cells and set dead load, see Chapter 5.14.4 or
-
only search for load cells, see Chapter 5.14.5
-
Assign load cells, see Chapter 5.14.6
-
Calibrate ‘New’: Max with weight unit, scale interval, dead load, calibration weight, see Chapter 5.14.7
-
Carry out a corner correction if necessary, see Chapter 5.14.10
Note
For further information about calibrating weighing points, see Chapter 5.11.
5.14.4 Search for load cells and set dead load
Access the menu as follows: [System setup]-[Weighing point]-[Weighing point A]-[Calib].
t
t
Unload the scale.
t
t
Select and confirm the first menu item.
t
y
A prompt window appears.
t
t
Press the [Continue] soft key to start the search
process.
Or
t
t
Press the [Cancel] soft key to return to the Pendeo
setup menu.
t
y
An input window appears.
t
t
Change the default value, if necessary and press
the [Ok] soft key, um die Eingabe zu bestätigen.
t
t
Press the [Cancel] soft key, if necessary to return
to the Pendeo setup menu.
t
y
Search process is started and dead load is set.