PR 5500 Instrument Manual
Commissioning
Sartorius EN-209
The empty scale (deck installed on a vehicle scale, no
load) for dead load:
t
t
Unload the scale, if necessary.
t
t
Press the [by load] soft key.
t
t
Confirm.
t
y
The confirmation is displayed with ‘setting dead
load...’.
Note
If the dead load is known, the value
can be overwritten by pressing [by
value].
t
t
Confirm the selection.
t
y
An input window appears.
t
t
Center the calibration weight on the scale and
enter the weight value.
t
t
Press the [OK] soft key.
t
t
Remove the calibration weight.
t
y
The corner load (dead load set to zero) is
displayed.
t
t
Carry out a corner correction if necessary, see
Chapter 5.14.10.
t
t
Press the [Save] soft key to save the calibration.