PR 5500 Instrument Manual
Device Installation
Sartorius EN-67
4.4.4.8 Connecting Weighing Platforms (CAP)
Up to 2 Combics analog platforms (CAP series) can be connected to the internal weighing electronics
connections depending on the model.
The cable colors shown above are valid for a CAPP4 500 x 400 and a CAPP1 320 x 420,
for example.
The assignments of cable colors are listed in the platform operating instructions.
!
Caution
The following example shows a platform with a 6-wire connection:
+ Meas.
– Meas.
+ Supply
+ Sense
– Sense
– Supply
equipotential bonding conductor
Potenzialausgleich
screen/Schirm
PR 5500
M+ M
–
V+ S
+
S
–
V
–
black/schwarz
blue/blau
green/grün
grey/grau
red/rot
white/weiß
or screen clamping rail/
oder Schirmklemmschiene
screen/
Schirm
Cable screens should be connected to the device earthing terminals or screen clamping rail.
Summary of Contents for Sartorius Intec Maxxis 4 PR 5500
Page 406: ......