Commissioning
PR 5500 Instrument Manual
EN-172 Sartorius
5.11.13 Saving the Calibration
t
t
Press the [Save] soft key.
t
y
This window appears for a short time.
The maximum capacity (Max) is displayed with the
ID ‘TST’ and without a weight unit.
t
y
The confirmation is displayed as follows:
‘saving calibration...’
t
t
After exiting calibration, go to the menu [Param]
and set the parameter ‘Settings locked’ to activate
overwrite protection via the software (see
Chapter 5.1.5.2).
Note
A unique check number is created every time a calibration or changed parameters are saved.
5.11.14 Canceling Calibration
t
t
Press the EXIT key to finish calibration.
t
y
You are prompted to confirm.
t
t
Press the [Yes] soft key to exit calibration without
saving.
t
y
Calibration is finished without saving with the
following display:
‘undoing calibration...’