Commissioning
PR 5500 Instrument Manual
EN-110 Sartorius
5.1.5.2 Via Software
Note
A unique check number is created every time a calibration or changed parameters are saved.
This can be viewed under [System information]-[Show calibration check numbers] (see
Chapter 5.20.7) and noted; if necessary.
The weighing point has a ‘Settings locked’ parameter
in the [Parameters] menu item.
If this parameter is set for the weighing point,
-
the calibration data and parameters of the
weighing point are only viewed as if the
corresponding CAL switch were closed.
-
Only the ‘Settings locked’ parameter can be
changed.
-
Restoring/importing a weighing point is locked.
Note
If ‘Settings locked’ and ‘W&M’ were selected for at least one weighing point, this has the
same effect as a closed CAL 1 and CAL 2 switch, see Chapter 5.1.5.1.