Device Installation
PR 5500 Instrument Manual
EN-48 Sartorius
4.4.1.4 Connecting a PR 5110 Remote Display via RS-485
Connection via the RS-485 interface, 4-wire transmission, point-to-point, full duplex (simultaneous sending and
receiving possible) with the PR 5110 remote display.
Example
PR 5500/04
RS-485-A
PR 5500
A-TxB
RxD/TxD-P
RxD/TxD-N
A-TxA
A-RxB
A-RxA
GNDI
+ 24 V
GND
DC
DC
1
2
3
4
5
6
7
8
↓
ON
A
{
S4
120
S1
S3
120
S2
1K6
1K6
A
RS-485
PR5110
4
5
3
2
1
D-Sub 9-contact male connector/
9-polige D-Sub Buchse (male)
5-pin plug-in terminal block/
5-polige steckbare Schraubklemme
2
8
3
7
5
RxB
RxA
TxB
TxA
RxD/TxD-P
RxD/TxD-N
GNDI
S4
S1
S2
S3
1K6
1K6
120
120
S5
DC
DC
1 2 3 4 5
ON
DIP
PR 5500 Switch settings
PR 5110 Switch settings
ON:
S1
ON:
S1, S2, S3
OFF:
S2, S3, S4
OFF:
S4, S5
PR 5500 Configuration
PR 5110 Configuration
[System setup] – [Connected devices] – [Remote display] –
[Interface] – [Option-1 RS-485-A]
–
–
–
45
–
2
–
–
–
3
–
–
–
4
–
WGH
–
W
–
5
–
WY
–
C
Summary of Contents for Sartorius Intec Maxxis 4 PR 5500