PR 5500 Instrument Manual
Device Installation
Sartorius EN-63
4.4.4.5 Connecting up to 8 Load Cells (650 ¥) using a 6-Wire Connecting Cable
Connections are made via junction box PR 6130/.. using connecting cable PR 6135.
The cable colors shown above are valid for Sartorius Intec load cells and connecting
cable PR 6135.
Color code:
bk
=
Black
rd
=
Red
bu
=
Blue
wh
=
White
gn
=
Green
ye
=
Yellow
gy
=
Gray
!
Caution
Recommendation
-
Install cable in steel pipe connected to a ground potential.
-
The distance between the measuring cables and the power cables should be at least 1 m.
Load cell supply circuit
-
Load resistance of load cell circuit
>
75 ¥, e.g. 8 load cells of 650 ¥ each
-
Supply voltage is 12 V DC, for further data, see Chapter 12.3.