PR 5500 Instrument Manual
Commissioning
Sartorius EN-221
t
y
After determining the Pendeo setup menu
appears.
5.15.6 View and Assign Load Cells
The load cells (serial number) can be assigned to the place of installation in this menu. This is important for
correcting the dead load (distribution to the individual load cells), for corner correction and in the event of load
cell replacement.
Note
The assignment from the installation should be documented in the case of load cells being
replaced.
Access the Pendeo setup menu as follows: [System setup]-[Weighing point] [Weighing point A]-[Calib]-[View].
t
t
Select and confirm the menu item ‘View and
assign load cells’.
5.15.6.1 View Load Cells
t
y
The load cells are displayed with their item
number, serial number, and load.
t
t
Select the desired line, if necessary and press the
[Info] soft key, to display the load cell data.
Note
If load cell names have been
assigned (see Chapter 5.18.9),
the view can be switched with
the soft key [by name].