Commissioning
PR 5500 Instrument Manual
EN-206 Sartorius
The load cells are assigned by placing minimum
weights on the scale.
t
t
Place the weight on the corner/load cell which
will be assigned no. 1.
t
y
As soon as the device detects the weight change,
the corresponding line is selected.
t
t
Press the [Accept] soft key.
t
y
The future LC no. appears at the far right of the
line.
t
t
Remove the calibration weight.
t
t
Repeat these steps for load cells 2–4.
t
t
Press the [Save] soft key to save the new
assignment.
1
2
3
4
1
2
3
4
5
6
7
8
1
2
3
4
5
6
1
2
3
4
5
6
7
8
An example of a possible assignment is shown on the
left.
t
y
The Pendeo setup menu appears.
t
t
Select and confirm the menu item ‘View and
assign load cells’.
Summary of Contents for Sartorius Intec Maxxis 4 PR 5500