Commissioning
PR 5500 Instrument Manual
EN-182 Sartorius
5.12 Configuring a xBPI Scale
5.12.1 General Information
The PR 5500 can communicate with scales (e.g., Combics1) or weighing modules via the xBPI protocol. The scale
can be connected via serial interface (PR 5500/04; see Chapter 4.4.1). Communication is serial.
Note
It’s possible to use the internal weighing point or an IS platform without external power
supply.
The determination and transmission of weight depends strongly on the scale/weighing module. Weight values up
to seven digits plus pre/- sign can be displayed. The maximum capacity (Max) must be set in the scale or
already set for a platform and cannot be changed via the xBPI protocol.
5.12.2 xBPI Parameter Input for Serial Interface
t
t
Go to [System setup]-[Weighing points] [Weighing
point A], select and confirm ‘xBPI scale’.
t
y
The following window appears.
t
t
Select and confirm the line ‘Interface’.
t
y
The following window appears.
t
t
Select and confirm the desired interface.
t
y
The previous window appears.
t
t
If necessary, change the parameters. Only the
‘Baudrate’ and ‘Stop bits’ can be set for an
xBPI scale.
t
t
Press the [Save] soft key to save the changes.