Velocity measurement problems and recommended actions
(continued)
Table 10-4:
Problem
Possible causes
Recommended actions
Erratic non-zero ve-
locity reading when
velocity is steady
•
Two-phase flow
•
Damping value too low
•
Plugged or coated sensor tube
•
Output wiring problem
•
Problem with receiving device
•
Wiring problem
•
Verify that the sensor orientation is appro-
priate for your application (refer to the
sensor installation manual).
•
Check the drive gain and the pickoff volt-
age. See
.
•
Check for air entrainment, tube fouling,
flashing, or tube damage.
•
Purge the sensor tubes.
•
Check for open or leaking valves or seals.
•
Check for sources of vibration.
•
Verify damping configuration.
•
Verify that the measurement units are con-
figured correctly for your application.
•
Check for two-phase flow. See
.
•
Check for radio frequency interference.
See
•
Contact Micro Motion.
Inaccurate velocity
reading
•
Wiring problem
•
Inappropriate measurement unit
•
Incorrect flow calibration factor
•
Incorrect density calibration factors
•
Incorrect grounding
•
Two-phase flow
•
Problem with receiving device
•
Incorrect sensor zero
•
Verify that the measurement units are con-
figured correctly for your application.
•
Verify all of the characterization or calibra-
tion parameters. See the sensor tag or the
calibration sheet for your meter.
•
.
•
Check the grounding of all components.
See
•
Check for two-phase flow. See
.
•
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
•
Replace transmitter.
Velocity switch trig-
gered erroneously
•
Incorrect sensor zero
•
.
Troubleshooting
154
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...