Load a concentration matrix using ProLink III
10.11
Velocity measurement problems
Velocity measurement problems and recommended actions
Table 10-4:
Problem
Possible causes
Recommended actions
Non-zero velocity
reading at no-flow
conditions or at zero
offset
•
Misaligned piping (especially in new instal-
lations)
•
Open or leaking valve
•
Incorrect sensor zero
•
.
•
Check for open or leaking valves or seals.
•
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
•
Contact Micro Motion.
Erratic non-zero ve-
locity reading at no-
flow conditions
•
Leaking valve or seal
•
Two-phase flow
•
Plugged or coated sensor tube
•
Incorrect sensor orientation
•
Wiring problem
•
Vibration in pipeline at rate close to sensor
tube frequency
•
Damping value too low
•
Mounting stress on sensor
•
Verify that the sensor orientation is appro-
priate for your application (refer to the
sensor installation manual).
•
Check the drive gain and the pickoff volt-
age. See
.
•
Purge the sensor tubes.
•
Check for open or leaking valves or seals.
•
Check for sources of vibration.
•
Verify damping configuration.
•
Verify that the measurement units are con-
figured correctly for your application.
•
Check for two-phase flow. See
.
•
Check for radio frequency interference.
See
•
Contact Micro Motion.
Troubleshooting
Configuration and Use Manual
153
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...