Option Description
Informa-
tional
Actions when fault is detected:
•
The alert is posted to the Alert List.
•
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
•
The status LED returns to green.
Status alerts and options for
Status Alert Severity
Status alerts and
Status Alert Severity
Table 5-1:
Alert number
Alert title
Default severity
User can reset severity
A001
EEPROM Error
Fault
No
A002
RAM Error
Fault
No
A003
No Sensor Response
Fault
Yes
A004
Temperature Overrange
Fault
No
A005
Velocity Overrange
Fault
Yes
A006
Characterization Required
Fault
Yes
A008
Density Overrange
Fault
Yes
A009
Transmitter Initializing/Warm-
ing Up
Ignore
Yes
A010
Calibration Failure
Fault
No
A011
Zero Calibration Failed: Low
Fault
Yes
A012
Zero Calibration Failed: High
Fault
Yes
A013
Zero Calibration Failed: Unsta-
ble
Fault
Yes
A014
Transmitter Failure
Fault
No
A016
Sensor Temperature (RTD)
Failure
Fault
Yes
A017
Sensor Case Temperature
(RTD) Failure
Informational
Yes
A020
Calibration Factors Missing
Fault
Yes
A021
Transmitter/Sensor/Software
Mismatch
Fault
No
A029
Internal Electronics Failure
Fault
No
A030
Incorrect Board Type
Fault
No
A033
Insufficient Pickoff Signal
Fault
Yes
A037
Sensor Check Failed
Fault
Yes
A038
Time Period Signal Out of
Range
Fault
No
Configure device options and preferences
80
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...