Options for
Discrete Output Polarity
Options for
Discrete Output Polarity
Table 6-5:
Polarity
Description
Active High
•
When asserted (condition tied to DO is true), the cir-
cuit draws as much current as it can, up to a maximum
of 10 mA.
•
When not asserted (condition tied to DO is false), the
circuit draws less than 1 mA.
Active Low
•
When asserted (condition tied to DO is true), the cir-
cuit draws less than 1 mA.
•
When not asserted (condition tied to DO is false), the
circuit draws as much current as it can, up to a maxi-
mum of 10 mA.
6.3.3
Configure
Discrete Output Fault Action
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Fault Processing
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Discrete Output
>
DO Fault Action
Overview
Discrete Output Fault Action
controls the behavior of the discrete output if the transmitter
encounters an internal fault condition.
Note
For some faults only: If
Fault Timeout
is set to a non-zero value, the transmitter will not implement the
fault action until the timeout has elapsed.
CAUTION!
Do not use
Discrete Output Fault Action
as a fault indicator. If you do, you may not be able to
distinguish a fault condition from a normal operating condition. If you want to use the discrete
output as a fault indicator, set
Discrete Output Source
to
Fault
and set
Discrete Output Fault Action
to
None
.
Procedure
Set
Discrete Output Fault Action
as desired.
The default setting is
None
.
Integrate the meter with the control system
92
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...