6.3
Configure the discrete output
The discrete output is used to report specific meter or process conditions. The discrete
output parameters control which condition is reported and how it is reported.
Depending on your purchase option, your transmitter may have one discrete output or no
discrete outputs.
Important
Whenever you change a discrete output parameter, verify all other discrete output parameters
before returning the meter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
•
(Section 6.3.1)
•
(Section 6.3.2)
•
(Section 6.3.3)
6.3.1
Configure
Discrete Output Source
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CH B
>
DO
>
CONFIG DO
>
DO SRC
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
Discrete Output
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Discrete Output
>
DO Source
Overview
Discrete Output Source
controls which device condition or process condition is reported via
the discrete output.
Procedure
Set
Discrete Output Source
to the desired option.
The default setting for
Discrete Output Source
is
Fault
.
Options for
Discrete Output Source
Options for
Discrete Output Source
Table 6-4:
Option
Label
State
Discrete output volt-
age
ProLink III
Field Communicator
Enhanced Event 1–5
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
ON
Site-specific
OFF
0 V
Calibration in Progress
Calibration in Progress
Calibration in Progress
ON
Site-specific
Integrate the meter with the control system
90
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...