API tables, process fluids, measurement units, and default reference values
Table 4-7:
Process fluid
API table
Referred density (API)
Default reference
temperature
Default reference
pressure
Generalized crude and JP4
5A
Unit: °API
Range: 0 to 100 °API
60 °F
0 psi (g)
23A
Unit: SGU
Range: 0.6110 to 1.0760
SGU
60 °F
0 psi (g)
53A
Unit: kg/m
3
Range: 610 to 1075 kg/m
³
15 °C
0 kPa (g)
Generalized products (gaso-
line, jet fuel, aviation fuel,
kerosene, heating oils, fuel
oils, diesel, gas oil)
5B
Unit: °API
Range: 0 to 85 °API
60 °F
0 psi (g)
23B
Unit: SGU
Range: 0.6535 to 1.0760
SGU
60 °F
0 psi (g)
53B
Unit: kg/m
3
Range: 653 to 1075 kg/m
³
15 °C
0 kPa (g)
Liquids with a constant den-
sity base or known thermal
expansion coefficient
6C
Unit: °API
60 °F
0 psi (g)
24C
Unit: SGU
60 °F
0 psi (g)
54C
Unit: kg/m
³
15 °C
0 kPa (g)
Lubricating oils
5D
Unit: °API
Range:
−
10 to +40 °API
60 °F
0 psi (g)
23D
Unit: SGU
Range: 0.8520 to 1.1640
SGU
60 °F
0 psi (g)
53D
Unit: kg/m
³
Range: 825 to 1164 kg/m
³
15 °C
0 kPa (g)
NGL (natural gas liquids)
23E
Unit: SGU
60 °F
0 psi (g)
LPG (liquid petroleum gas)
24E
Unit: SGU
60 °F
0 psi (g)
Restriction
These tables are not appropriate for the following process fluids: propane and propane mixes,
butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, ethylene, propylene,
cyclohexane, aeromatics, asphalts, and road tars.
Set up temperature and pressure data for API referral using
ProLink III
The API referral application uses temperature and pressure data in its calculations. You
must decide how to provide this data, then perform the required configuration and setup.
Configure process measurement
Configuration and Use Manual
47
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...