density. This value is used to calculate specific gravity. The result of the specific gravity
calculation is then used in the equations used to calculate °Baumé, °Brix, °Plato, or
°Twaddell.
Specific gravity is always calculated using the two reference temperatures that are
specified during concentration measurement configuration.
If you are measuring in °Baumé, the meter will automatically select the appropriate
equation according to the specific gravity of the process fluid, and will switch equations
when specific gravity crosses 1.0.
Equations used for specific gravity, °Baumé, °Brix, °Plato, and °Twaddell
Table 4-9:
Equation
Current value of
specific gravity
Equation
Specific Gravity
N/A
SG =
ρ
RefTemp
ρ
WaterRefTemp
°Baumé
1.0 or greater
(°Baumé heavy)
°Baumé = 145 −
145
SG
Less than 1.0
(°Baumé light)
°Baumé =
140
SG
− 130
°Brix
N/A
°Brix = 318.906 −
(
384.341
SG
)
+
(
66.1086
SG2
)
°Plato
N/A
°Plato = (668.72 × SG) − 463.37 −
(
205.347 × SG
2
)
°Twaddell
N/A
°Twaddell = 200 × (SG − 1)
SG
Specific gravity of process fluid
ρ
RefTemp
Density of process fluid at
Reference Temperature for Referred Density
, as measured
using the active concentration matrix
ρ
WaterRefTemp
Density of water at
Reference Temperature for Water
Related information
Set reference temperature values for specific gravity using ProLink III
Set reference temperature values for specific gravity using the Field Communicator
4.7.5
Matrix switching
Matrix switching can be used to measure different process fluids without resetting the
active matrix manually. Matrix switching can also be used to increase measurement
accuracy.
When matrix switching is enabled, the meter automatically switches between the matrices
in Slot 1 and Slot 2 whenever an extrapolation alert is present for the active matrix but
would not be generated by the other matrix. For example:
Configure process measurement
Configuration and Use Manual
71
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...