For these terms:
•
You can specify a constant value.
•
You can specify a process variable. If you choose this, the current value of the
process variable is used in the calculation.
Important
User-defined calculations are performed using the meter's internal measurement units.
Therefore:
•
If a constant represents a process variable, you must enter its value in the internal
measurement units.
•
If a constant will be used to modify a process variable, you must use the internal
measurement units to derive the constant.
3.
If you selected User-Defined Calculation 2:
a. Enter the values to be used for the constants: A, B, C.
b. Enter the value to be used for
t
.
For this term:
•
You can specify a constant value.
•
You can specify a process variable. If you choose this, the current value of the
process variable is used in the calculation.
Important
User-defined calculations are performed using the meter's internal measurement units.
Therefore:
•
If a constant represents a process variable, you must enter its value in the internal
measurement units.
•
If a constant will be used to modify a process variable, you must use the internal
measurement units to derive the constant.
4.
If you selected User-Defined Calculation 3:
a. Enter the values to be used for the constants: A, B, C, E, F.
b. Enter the value to be used for
⍴
W
(the density of water at reference temperature
and reference pressure)
Restriction
User-Defined Calculation 3 is available only if the concentration measurement application is
enabled and a matrix is active.
Important
User-defined calculations are performed using the meter's internal measurement units.
Therefore:
•
If a constant represents a process variable, you must enter its value in the internal
measurement units.
Measurement support
132
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...