4.4
Configure the pressure input
Pressure data is required for several different measurements. The meter does not measure
pressure. There are several different methods to obtain pressure data.
•
Configure the pressure input using ProLink III
(Section 4.4.1)
•
Configure the pressure input using the Field Communicator
(Section 4.4.2)
4.4.1
Configure the pressure input using ProLink III
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Line Pressure
Overview
Pressure data is required for several different measurements. There are several different
methods to obtain pressure data.
Tip
A fixed pressure value is not recommended. A fixed pressure value may produce inaccurate process
data.
Prerequisites
You must be able to supply pressure data to the meter. The meter does not measure
pressure.
You must be using gauge pressure.
If you plan to poll an external device, the primary mA output must be wired to support
HART communications.
Procedure
1.
Choose
Device Tools
>
Configuration
>
Process Measurement
>
Line Pressure
.
2.
Set
Pressure Unit
to the unit used by the external pressure device.
Restriction
If the API referral application is enabled, the API table selection automatically sets the
pressure measurement unit. Configure the API referral application first, then change the
pressure measurement unit if necessary.
3.
Choose the method you will use to supply pressure data, and perform the required
setup.
Configure process measurement
34
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...