Temperature measurement problems and recommended actions
(continued)
Table 10-3:
Problem
Possible causes
Recommended actions
Inaccurate tempera-
ture data from exter-
nal device
•
Wiring problem
•
Problem with input configuration
•
Problem with external device
•
Problem with input configuration
•
Verify the wiring between the transmitter
and the external device.
•
Verify that the external device is operating
correctly.
•
Verify the configuration of the tempera-
ture input.
•
Ensure that both devices are using the
same measurement unit.
Related information
10.8.1
Thermal insulation
Temperature measurement is a potential source of significant error in density
measurement. Thermal insulation helps to maintain a constant temperature, and can
reduce or eliminate temperature problems.
Thermal insulation is especially important when the line temperature and the ambient
temperature are significantly different, or when there are abrupt changes in either the line
temperature or the ambient temperature.
A thermal insulation jacket should protect both the sensor body and the flanges.
A thermal insulation jacket designed specifically for your meter is available from
Micro Motion. See the product data sheet for your meter.
10.9
API referral problems
Problem
Possible causes
Recommended actions
No referred density
reading
•
Line pressure, line temperature, or line
density is outside the range of the config-
ured API table
•
Check process conditions.
•
Ensure that the API referral application is
configured correctly for your process fluid.
See
Troubleshooting
Configuration and Use Manual
151
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...