Tip
The lower the precision, the greater the change must be for it to be reflected on the display.
Do not set the precision too low or too high to be useful.
5.1.4
Configure the refresh rate of data shown on the display
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
DISPLAY
>
RATE
ProLink III
Device Tools
>
Configuration
>
Transmitter Display
>
Display Variables
Field Communicator
Configure
>
Manual Setup
>
Display
>
Display Behavior
>
Refresh Rate
Overview
You can set
Refresh Rate
to control how frequently data is refreshed on the display.
Procedure
Set
Refresh Rate
to the desired value.
The default value is 1000 milliseconds. The range is 100 milliseconds to
10,000 milliseconds (10 seconds).
5.1.5
Enable or disable automatic scrolling through the
display variables
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
DISPLAY
>
AUTO SCRLL
ProLink III
Device Tools
>
Configuration
>
Transmitter Display
>
General
Field Communicator
Configure
>
Manual Setup
>
Display
>
Display Behavior
>
Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display
variables or to show a single display variable until the operator activates
Scroll
. When you
set automatic scrolling, you can also configure the length of time each display variable is
displayed.
Procedure
1.
Enable or disable
Auto Scroll
as desired.
Option
Description
Enabled
The display automatically scrolls through each display variable as specified
by
Scroll Rate
. The operator can move to the next display variable at any
time using
Scroll
.
Configure device options and preferences
Configuration and Use Manual
75
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...