Procedure
1.
Choose
Configure
>
Manual Setup
>
Measurements
>
API Referral
.
2.
Choose
API Referral Setup
.
3.
Specify the API table that you want to use for measurement.
Each API table is associated with a specific set of equations.
a. Set
API Table Number
to the number that matches the API table units that you
want to use for referred density.
Your choice also determines the measurement unit to be used for temperature
and pressure, and the default values for reference temperature and reference
pressure.
API Table Number
Measurement
unit for referred
density
Temperature
measurement
unit
Pressure meas-
urement unit
Default refer-
ence tempera-
ture
Default refer-
ence pressure
5
°API
°F
psi (g)
60 °F
0 psi (g)
6
°API
°F
psi (g)
60 °F
0 psi (g)
23
SGU
°F
psi (g)
60 °F
0 psi (g)
SGU
°F
psi (g)
60 °F
0 psi (g)
53
kg/m
³
°C
kPa (g)
15 °C
0 kPa (g)
kg/m
³
°C
kPa (g)
15 °C
0 kPa (g)
(2) Used only with
API Table Letter
=
C
.
b. Set
API Table Letter
to the letter of the API table group that is appropriate for your
process fluid.
API Table Letter
Process fluids
A
Generalized crude and JP4
B
Generalized products: Gasoline, jet fuel, aviation fuel, kerosene,
heating oils, fuel oils, diesel, gas oil
C
(3)
Liquids with a constant base density or known thermal expansion
coefficient (TEC). You will be required to enter the TEC for your
process fluid.
D
Lubricating oils
Configure process measurement
Configuration and Use Manual
51
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...