Related information
Fault indication with the discrete output
Options for
Discrete Output Fault Action
Options for
Discrete Output Fault Action
Table 6-6:
Label
Discrete output behavior
Polarity=
Active High
Polarity=
Active Low
Upscale
•
Fault: discrete output is ON
(site-specific voltage)
•
No fault: discrete output is con-
trolled by its assignment
•
Fault: discrete output is OFF
(0 V)
•
No fault: discrete output is con-
trolled by its assignment
Downscale
•
Fault: discrete output is OFF
(0 V)
•
No fault: discrete output is con-
trolled by its assignment
•
Fault: discrete output is ON
(site-specific voltage)
•
No fault: discrete output is con-
trolled by its assignment
None
(default)
Discrete output is controlled by its assignment
Fault indication with the discrete output
To indicate faults via the discrete output, set
Discrete Output Source
to
Fault
. Then, if a fault
occurs, the discrete output is always ON and the setting of
Discrete Output Fault Action
is
ignored.
6.4
Configure an enhanced event
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Events
>
Enhanced Events
Field Communicator
Configure
>
Alert Setup
>
Enhanced Events
Overview
An enhanced event is used to provide notification of process changes. An enhanced event
occurs (is ON) if the real-time value of a user-specified process variable moves above (HI)
or below (LO) a user-defined setpoint, or in range (IN) or out of range (OUT) with respect
to two user-defined setpoints. You can define up to five enhanced events.
Procedure
1.
Select the event that you want to configure.
2.
Specify
Event Type
.
Integrate the meter with the control system
Configuration and Use Manual
93
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...