10.4.3
Perform loop tests using the Field Communicator
A loop test is a way to verify that the transmitter and the remote device are
communicating properly. A loop test also helps you know whether you need to trim mA
outputs.
Prerequisites
Before performing a loop test, configure the channels for the transmitter inputs and
outputs that will be used in your application.
Follow appropriate procedures to ensure that loop testing will not interfere with existing
measurement and control loops.
Procedure
1.
Test the mA output(s).
a. Choose
Service Tools
>
Simulate
>
Simulate Outputs
>
mA Output 1 Loop Test
or
Service
Tools
>
Maintenance
>
Simulate Outputs
>
mA Output 2 Loop Test
, and select
4 mA
.
b. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
c. Press
OK
.
d. Select
20 mA
.
e. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
f. Press
OK
.
g. Choose
End
.
2.
Test the discrete output(s).
a. Press
Service Tools
>
Simulate
>
Simulate Outputs
>
Discrete Output Test
.
b. Choose
Off
.
c. Verify the signal at the receiving device.
d. Press
OK
.
e. Choose
On
.
f. Verify the signal at the receiving device.
g. Press
OK
.
h. Choose
End
.
3.
Test the TPS output.
Troubleshooting
Configuration and Use Manual
141
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...