Sample calibration certificate: K format, SI units
Figure A-2:
DENSITY CALIBRATION CERTIFICATE
CDM100PA18MBAZ00EANZZ
2012.10031
Model Code:
Customer Tag:
Serial No:
Calibration Date:
Pressure Test:
2013-07-01 11:05:22
225 bar(g)
Calibration ID:
1.29825731
7835|7845 "K" Coefficients (SI Units)
K0 =
-3.000
Density Calibration Coefficients @20°C
Temperature Compensation Coefficients:
Pressure Compensation Coefficients:
K18 =
-3.154066E-004
K19 =
-9.992198E-004
K20A=
1.238665E-004
Calibrated Range = 300-1300 kg/m³ (0.3-1.300 g/cm³)
K1 =
0
K2 =
1.432987E-007
K20B=
K21A=
K21B=
Range: 1-41 bar(a)
Range: 31-71 bar(a)
Range: 61-101 bar(a)
Range: 101-151 bar(a)
1.238665E-004
1.238665E-004
1.238665E-004
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Time Period, TP sec
μ
Density (kg/m³)
Density Calibration Data:
5065.2054
0.97
5587.3151
798.82
5857.0031
1240.79
Performance Check:
Error (kg/m³)
Density (kg/m³)
998.22
-0.08
Known Density Verification Data:
765.1
Verification Time Period (Air) @ 20°C =
μsec
CALIBRATED BY:
D = Density (uncompensated)
TP = Time period
P= Pressure
Definitions:
°C
kg/m³
kg/m³
Dt = Density (Temperature compensated)
Dp = Density (Pressure & Temperature compensated)
kg/m³
µsec
T = Temperature
bar(a)
Form Rev: 1
2013-08-29 13:55:05
Printed
Page 2 of 3
All equipment used for this calibration is calibrated at routine intervals against
standards that are traceable to the International System of Units (SI).
Emerson Process Management Micro Motion, Inc. 7070 Winchester Circle, Boulder, CO 80301
SAMPLE
Calibration certificate
Configuration and Use Manual
169
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...