Procedure
1.
Collect data for Comparison 1.
a. Take a concentration reading from the meter and record line density and line
temperature.
b. Take a sample of the process fluid at the current concentration.
c. Obtain a laboratory value for concentration at line density and line temperature,
in the units used by the meter.
2.
Collect data for Comparison 2.
a. Change the concentration of your process fluid.
b. Take a concentration reading from the meter and record line density and line
temperature.
c. Take a sample of the process fluid at the current concentration.
d. Obtain a laboratory value for concentration at line density and line temperature,
in the units used by the meter.
3.
Populate the following equation with values from each comparison.
Concentration
Lab
=
(
A
× Concentration
Meter
)
+
B
4.
Solve for A (slope).
5.
Solve for B (offset), using the calculated slope and one set of values.
6.
Enter the results as the trim slope and the trim offset.
•
Using the display: Not available
•
Using ProLink III:
Device Tools
>
Configuration
>
Process Measurement
>
Concentration
Measurement
•
Using the Field Communicator:
Configure
>
Manual Setup
>
Measurements
>
Optional
Setup
>
Concentration Measurement
>
Trim CM Process Variables
7.
Take another concentration reading from the meter, and compare it to the
laboratory value.
•
If the two values are acceptably close, the trim is complete.
•
If the two values are not acceptably close, repeat this procedure.
Example: Calculating the trim offset and the trim slope
Comparison 1
Laboratory value
50.00%
Meter value
49.98%
Comparison 2
Laboratory value
16.00%
Meter value
15.99%
Populate the equations:
50 = (
A
× 49.98) +
B
Measurement support
130
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...