4.5.2
Configure
Velocity Damping
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Velocity
>
Velocity Damping
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Velocity
>
Velocity Damping
Overview
Velocity Damping
controls the amount of damping that is applied to the velocity
measurement.
Damping is used to smooth out small, rapid fluctuations in process measurement.
Damping
Value
specifies the time period (in seconds) over which the transmitter will spread changes
in the process variable. At the end of the interval, the internal value will reflect 63% of the
change in the actual measured value.
Procedure
Set
Velocity Damping
to the value you want to use.
The default value is 0.8 seconds. The range is 0.0–51.2 seconds.
The value is automatically rounded down to the nearest valid value. Valid values are as
follows: 0, 0.2, 0.4, 0.8, … 51.2.
4.5.3
Configure
Flow Direction
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Velocity
>
Flow Direction
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Velocity
>
Flow Direction
Overview
Flow Direction
controls how forward flow and reverse flow affect velocity measurement and
reporting.
Flow Direction
is defined with respect to the flow arrow on the sensor:
•
Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
•
Reverse flow (negative flow) moves in the direction opposite to the flow arrow on
the sensor.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow
direction or the setting of the
Flow Direction
parameter.
Configure process measurement
Configuration and Use Manual
39
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...