Option
Description
Disabled
(de-
fault)
The display shows
Display Variable 1
and does not scroll automatically. The
operator can move to the next display variable at any time using
Scroll
.
2.
If you enabled
Auto Scroll
, set
Scroll Rate
as desired.
The default value is 10 seconds.
Tip
Scroll Rate
may not be available until you apply
Auto Scroll
.
5.2
Enable or disable operator actions from the
display
You can configure the transmitter to let the operator perform specific actions using the
display.
•
(Section 5.2.1)
5.2.1
Enable or disable the
Acknowledge All Alerts
display
command
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
DISPLAY
>
ACK
ProLink III
Device Tools
>
Configuration
>
Transmitter Display
>
Ack All
Field Communicator
Configure
>
Manual Setup
>
Display
>
Display Menus
>
Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge
all alerts from the display.
Procedure
1.
Ensure that the alert menu is accessible from the display.
To acknowledge alerts from the display, operators must have access to the alert
menu.
2.
Enable or disable
Acknowledge All Alerts
as desired.
Option
Description
Enabled
(default)
Operators can use a single display command to acknowledge all alerts at
once.
Configure device options and preferences
76
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...