If the fault timeout period expires while the alert is still active, the fault actions are
performed. If the alert condition clears before the fault timeout expires, no fault actions
are performed.
5.4.2
Configure
Status Alert Severity
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Alert Severity
Field Communicator
Configure
>
Alert Setup
>
Alert Severity
>
Set Alert Severity
Overview
Use
Status Alert Severity
to control the fault actions that the transmitter performs when it
detects an alert condition.
Restrictions
•
For some alerts,
Status Alert Severity
is not configurable.
•
For some alerts,
Status Alert Severity
can be set only to two of the three options.
Tip
Micro Motion recommends using the default settings for
Status Alert Severity
unless you have a specific
requirement to change them.
Procedure
1.
Select a status alert.
2.
For the selected status alert, set
Status Alert Severity
as desired.
Option Description
Fault
Actions when fault is detected:
•
The alert is posted to the Alert List.
•
Outputs go to the configured fault action (after
Fault Timeout
has expired, if ap-
plicable).
•
Digital communications go to the configured fault action (after
Fault Timeout
has expired, if applicable).
•
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
•
Outputs return to normal behavior.
•
Digital communications return to normal behavior.
•
The status LED returns to green.
Configure device options and preferences
Configuration and Use Manual
79
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...