Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the
transmitter can store up to six concentration matrices, only one matrix can be used for
measurement at any one time.
1.
Choose
Device Tools
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
2.
Scroll to Step 2, set
Active Matrix
to the matrix you want to use and click
Change Matrix
.
Note
To support matrix switching, you must select the matrix in Slot 1 or the matrix in Slot 2.
Related information
4.7.3
Set up concentration measurement using the
Field Communicator
This section guides you through most of the tasks related to setting up and implementing
the concentration measurement application.
Restrictions
•
This section does not cover building a concentration matrix. See
Micro Motion Enhanced
Density Application: Theory, Configuration, and Use
for detailed information on building a
matrix.
•
You cannot load a concentration matrix using the Field Communicator. If you need to load a
matrix, you must use ProLink III.
1.
Enable the concentration measurement application using the Field Communicator
2.
Set reference temperature values for specific gravity using the Field Communicator
3.
Provide temperature data for concentration measurement using the Field Communicator
4.
Modify matrix names and labels using the Field Communicator
5.
Modify concentration measurement operational parameters using the
Field Communicator
6.
Select the active concentration matrix using the Field Communicator
Enable the concentration measurement application using
the Field Communicator
The concentration measurement application must be enabled before you can perform any
setup. If the concentration measurement application was enabled at the factory, you do
not need to enable it now.
1.
Choose
Overview
>
Device Information
>
Applications
>
Enable/Disable Applications
.
Configure process measurement
Configuration and Use Manual
65
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...