Option
Description
Enabled
Operator is prompted for the off-line password at entry to the off-line
menu.
Disabled
(default)
No password is required for entry to the off-line menu.
4.
Set
Off-Line Password
to the desired value.
The default value is 1234. The range is 0000 to 9999.
Tip
Record your password for future reference.
5.4
Configure alert handling
The alert handling parameters control the transmitter’s response to process and device
conditions.
•
(Section 5.4.1)
•
(Section 5.4.2)
5.4.1
Configure
Fault Timeout
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Fault Processing
Field Communicator
Configure
>
Alert Setup
>
Alert Severity
>
Fault Timeout
Overview
Fault Timeout
controls the delay before fault actions are performed.
Restriction
Fault Timeout
is applied only to the following alerts (listed by Status Alert Code): A003, A004, A005,
A008, A016, A017, A033. For all other alerts, fault actions are performed as soon as the alert is
detected.
Procedure
Set
Fault Timeout
as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set
Fault Timeout
to 0, fault actions are performed as soon as the alert condition is
detected.
The fault timeout period begins when the transmitter detects an alert condition. During
the fault timeout period, the transmitter continues to report its last valid measurements.
Configure device options and preferences
78
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...