Verify that the host has access to the required data.
-
Verify that the host is writing to the correct register in memory, using the correct data
type.
•
If necessary, apply an offset.
Modify matrix names and labels using ProLink III
For convenience, you can change the name of a concentration matrix and the label used
for its measurement unit. This does not affect measurement.
1.
Choose
Device Tools
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
2.
Scroll to Step 2, set
Matrix Being Configured
to the matrix you want to modify, and click
Change Matrix
.
3.
Scroll to Step 3, then perform the following actions:
a. Set
Concentration Units Label
to the label that will be used for the concentration
unit.
b. If you set
Concentration Units Label
to
Special
, enter the custom label in
User-Defined
Label
.
c. In
Matrix Name
, enter the name to be used for the matrix.
4.
Click the
Apply
button at the bottom of Step 3.
Modify operational parameters for concentration
measurement using ProLink III
You can enable and disable extrapolation alerts, set extrapolation alert limits, and control
matrix switching. These parameters control the behavior of the concentration
measurement application but do not affect measurement directly. Additionally, for certain
types of concentration measurement, you can select the calculation method to be used.
1.
Choose
Device Tools
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
2.
Scroll to Step 2, set
Matrix Being Configured
to the matrix you want to modify, and click
Change Matrix
.
3.
Scroll to Step 4.
4.
Set up extrapolation alerts.
Each concentration matrix is built for a specific density range and a specific
temperature range. If line density or line temperature goes outside the range, the
transmitter will extrapolate concentration values. However, extrapolation may
Configure process measurement
Configuration and Use Manual
63
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...