Standard concentration matrices and associated measurement units
(continued)
Table E-1:
Matrix name
Description
Density unit
Temperature
unit
Derived variable
HFCS 42
Matrix represents a hydrometer scale
for HFCS 42 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
g/cm
3
°C
Mass Concentration
(Density)
HFCS 55
Matrix represents a hydrometer scale
for HFCS 55 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
g/cm
3
°C
Mass Concentration
(Density)
HFCS 90
Matrix represents a hydrometer scale
for HFCS 90 (high-fructose corn syrup)
solutions that indicates the percent by
mass of HFCS in solution.
g/cm
3
°C
Mass Concentration
(Density)
E.2
Concentration measurement matrices
available by order
Depending on your product, you may be able to order one or more of the matrices listed
here. You can have it loaded at the factory or you can load it from a file.
Concentration matrices, names, ranges, units, and derived variable
Table E-2:
Process fluid
Matrix file name
Default ma-
trix name
Concentra-
tion range
Tempera-
ture range
Density
unit
Tem-
pera-
ture
unit
Derived var-
iable
Alcohol (by vol-
ume)
Alcohol 0–20%Vol
0–40C.xml
Alcohol
0–20% (by
volume)
0–40 °C
g/cm
³
°C
Concentration
(Density)
Alcohol 50–
100%Vol 40–
70C.xml
Alcohol
50–100% (by
volume)
40–70 °C
g/cm
³
°C
Concentration
(Density)
Alcohol 80–
100%Vol 50–
90C.xml
Alcohol
80–100% (by
volume)
50–90 °C
g/cm
³
°C
Concentration
(Density)
Alcohol (proof)
Alcohol 100–200
Proof 50–70C.xml
Alcohol
100–200
(proof)
50–70 °C
g/cm
³
°C
Concentration
(Density)
Alcohol 160–200
Proof 50–90C.xml
Alcohol
160–200
(proof)
50–90 °C
g/cm
³
°C
Concentration
(Density)
Sucrose solution
(Brix)
Sucrose 0–40 Brix
0–100C.xml
Sucrose
0–40%
0–100 °C
g/cm
³
°C
Concentration
(Density)
Concentration measurement matrices, derived variables, and process variables
208
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...