Alert num-
ber
Alert title
Possible cause
Recommended actions
A102
Drive Overrange
The drive power (current/voltage) is
at its maximum.
•
Check the drive gain and the pickoff
voltage. See
and
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Check for fluid separation by monitor-
ing the density value and comparing
the results against expected density val-
ues.
•
Ensure that the sensor orientation is ap-
propriate for your application. Settling
from a two-phase or three-phase fluid
can cause this alert.
A104
Calibration in Pro-
gress
A calibration procedure is in proc-
ess.
•
Allow the procedure to complete.
•
For zero calibration, you may abort the
calibration, set
Zero Time
to a lower val-
ue, and restart the calibration.
A105
Two-Phase Flow
The line density is outside the user-
defined two-phase flow limits.
•
Check for two-phase flow. See
A106
Burst Mode Enabled
HART burst mode is enabled.
•
No action required.
•
If desired, you can set
Alert Severity Level
to
Ignore
.
A107
Power Reset Occur-
red
The transmitter has been restarted.
•
No action required.
•
If desired, you can set
Alert Severity Level
to
Ignore
.
A113
mA Output 2 Satura-
ted
The calculated mA output value is
outside the configured range.
•
Check the settings of
Upper Range Value
and
Lower Range Value
. See
•
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Verify that the measurement units are
configured correctly for your applica-
tion.
A114
mA Output 2 Fixed
The mA output is configured to
send a constant value.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Exit mA output trim, if applicable.
•
Check whether the output has been set
to a constant value via digital communi-
cation.
Troubleshooting
Configuration and Use Manual
147
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...