Possible causes and recommended actions for excessive (saturated)
drive gain
(continued)
Table 10-7:
Possible cause
Recommended actions
Flow rate out of range
Ensure that the flow rate is within sensor limits.
Incorrect sensor characteriza-
tion
Verify the characterization or calibration parameters.
Erratic drive gain
Possible causes and recommended actions for erratic drive gain
Table 10-8:
Possible cause
Recommended actions
Two-phase flow
Check for two-phase flow. See
.
Foreign material caught in sen-
sor tubes
•
Purge the sensor tubes.
•
Replace the sensor.
10.24.1
Collect drive gain data
Drive gain data can be used to diagnose a variety of process and equipment conditions.
Collect drive gain data from a period of normal operation, and use this data as a baseline
for troubleshooting.
Procedure
1.
Navigate to the drive gain data.
2.
Observe and record drive gain data over an appropriate period of time, under a
variety of process conditions.
10.25
Check the pickoff voltage
If the pickoff voltage readings are unusually low, you may have any of a variety of process
or equipment problems.
To know whether your pickoff voltage is unusually low, you must collect pickoff voltage
data during the problem condition and compare it to pickoff voltage data from a period of
normal operation.
Possible causes and recommended actions for low pickoff voltage
Table 10-9:
Possible cause
Recommended actions
Cavitation or flashing; settling of
two-phase or three-phase fluids
•
Increase the inlet or back pressure at the sensor.
•
If a pump is located upstream from the sensor, increase
the distance between the pump and sensor.
•
The sensor may need to be reoriented or repositioned.
Consult the installation manual for your sensor.
Troubleshooting
164
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...