Default settings for
mA Output Process Variable
Table 6-1:
Device
Channel
mA output
Default process variable assign-
ment
CDM mA
Channel A
Primary mA output
Density
Channel B
Secondary mA output
Temperature
CDM DO
Channel A
Primary mA output
Density
CDM TPS
Channel A
Primary mA output
Temperature
Postrequisites
If you changed the setting of
mA Output Process Variable
, verify the settings of
Lower Range
Value
(LRV) and
Upper Range Value
(URV).
Options for
mA Output Process Variable
The transmitter provides a basic set of options for
mA Output Process Variable
, plus several
application-specific options. Different communications tools may use different labels for
the options.
Options for
mA Output Process Variable
Table 6-2:
Process variable
Label
Display
ProLink III
Field Communicator
Standard
Line Density
DENS
Line Density
Density
Line Temperature
TEMP
Line Temperature
Temperature
Case Temperature
TCASE
Case Temperature
Case Temperature
Tube-Case Temperature Dif-
ference
TDIFF
Tube-Case Temperature Dif-
ference
Tube-Case Temp Difference
Line Temperature (External
or Fixed)
EXT T
Line Temperature (External
or Fixed)
External Temperature
Line Pressure (External or
Fixed)
EXTP
Line Pressure (External or
Fixed)
External Pressure
Drive Gain
DGAIN
Drive Gain
Drive Gain
Sensor Time Period
TP B
Sensor Time Period
Sensor Time Period
Velocity
VEL
Velocity
Velocity
User-Defined Calculation
Output
UCALC
User-Defined Calculation
Output
User-Defined Calculation
Output
API referral
Referred Density (API)
RDENS
Referred Density (API)
Density at Reference (API)
Integrate the meter with the control system
Configuration and Use Manual
85
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...