AD-4412-CW 

Weighing Indicator 

 

CONNECTION SETTING EXAMPLE 

1WMPD4004155 

Summary of Contents for AD-4412-CW

Page 1: ...AD 4412 CW Weighing Indicator CONNECTION SETTING EXAMPLE 1WMPD4004155...

Page 2: ...duced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the i...

Page 3: ...detect basis Setting the product length Refer to 3 Installing the Weighing Indicator for details of connection Refer to 11 Interface for other external equipments Refer to 6 1 Calibrating the Indicato...

Page 4: ...struction manual 2 When using the photo eye sensor connect it to the I F terminal on the rear panel of the indicator Refer to Connection of Photo eye Sensor of 11 Interface in the instruction manual I...

Page 5: ...icator is not grounded it may cause an electric shock or malfunction due to discharge of static electricity Live terminal Neutral terminal 100 240 V 50Hz or 60Hz Noiseless stable single phase power so...

Page 6: ...l voltage 2V max Output circuit DO method Open collector Output circuit withstand voltage DC 40 V Max drive current 50 mA Output terminal residual voltage 1 5 V at drive current 50 mA Connecting the c...

Page 7: ...theinstructionmanualfordetails Table 3 Correspondence table of OP 02 and DO Input terminal DI number Output terminal DO number A1 DI 1 B1 DO 1 A2 DI 2 B2 DO 2 A3 DI 3 B3 DO 3 A4 DI 4 B4 DO 4 A5 DI 5 B...

Page 8: ...lot2 Output terminal DOnumber optionalslot1 DOnumber optionalslot2 A1 DI 12 DI 28 B1 DO 12 DO 28 A2 DI 13 DI 29 B2 DO 13 DO 29 A3 DI 14 DI 30 B3 DO 14 DO 30 A4 DI 15 DI 31 B4 DO 15 DO 31 A13 DI 24 DI...

Page 9: ...for the indicator The following settings are described based on this example Othrer specifications Scale weighing capacity 1000 g Conveyer length 300 mm Conveyer speed 80 m min Relay Conveyer Fig 8 Co...

Page 10: ...require quality manager or higher permissions Refer to 5 5 Managing User Level for Logging In of the instruction manual for setting the management level 2 1 Weighing unit decimal point position and di...

Page 11: ...the setting screen 2 Touch the Weighing button on the Balance tab and enter an appropriate capacity rated capacity for your load cell into the field Capacity on the Weighing Setting 1 tab Example belo...

Page 12: ...g the weight 1 Touch the Setting key on the weighing screen to display the setting screen 2 Touch the Cal button on the Balance tab to display the weight adjustment screen Fig 14 Weight adjustment scr...

Page 13: ...layed after finishing zero point calibration Enter a mass for the weight to be used for calibration The weight value can be set between 0 g and capacity g of the load cell 5 The Span calibration scree...

Page 14: ...in the memory even after shutting down the power of the indicator until the indicator is initialized or those settings are changed The following settings require quality manager or higher permissions...

Page 15: ...AD4412 CW 15 3 Set the length for your conveyer in the field Conveyer Length on the Main Unit 1 tab The indicator must be restarted to enable those settings Fig 20 Setting the conveyer length...

Page 16: ...uch the Setting key on weighing screen to display the setting screen Touch the Main Unit button on System 1 tab to display main unit setting screen 2 Set the Curb Chattering Photo Eye Sensor Polarity...

Page 17: ...tained in the memory even after shutting down the power of the indicator until the indicator is initialized or those settings are changed The following settings require quality manager or higher permi...

Page 18: ...play the weighing parameter setting screen 4 Enter an appropriate speed for your belt conveyor in the field Speed on the Weighing Parameter Setting 1 tab The speed for the belt conveyor must be calibr...

Page 19: ...the weighing parameter setting screen 2 Set an appropriate length for products to be conveyed in the field Product Length on the Weighing Parameter Setting 1 tab The product length is used for contin...

Page 20: ...the lower limit as being underweight When setting the reject stage to stage 5 the upper upper limit Hi Hi Limit value detects very overweight and lower lower limit Lo Lo Limit detects very underweight...

Page 21: ...output to the tower light is set to DO2 DO number is a number set by yourself at connection Setting the DO 1 Touch the Product key on the weighing screen to display the product screen Then touch the E...

Page 22: ...DO Map button to display the DO map screen 4 The example is when is placed in the field Belt Running for DO1 to operate the belt conveyer while weighing is placed in the field HiHi and LoLo for DO2 to...

Page 23: ...detecting it at the end of the conveyer plus the swing time for the flipper of the rejecter However the swing time for the flipper of the rejecter varies depending on settings for the cushion needle...

Page 24: ...ptions are only for basic connection methods and functions the indicator can also be practially used by adding functions or options not described in this document such as providing feedback to the pac...