Added Damping
affects the reporting of
mA Output Process Variable
through the mA output
only. It does not affect the reporting of that process variable via any other method (e.g., a
frequency output or digital communications), or the value of the process variable used in
calculations.
Note
Added Damping
is not applied if the mA output is fixed (for example, during loop testing) or if the mA
output is reporting a fault.
Added Damping
is applied while sensor simulation is active.
Procedure
Set
Added Damping
to the desired value.
The default value is 0.0 seconds. The range is 0.0 to 440 seconds.
When you specify a value for
Added Damping
, the transmitter automatically rounds the value
down to the nearest valid value.
Interaction between
mA Output Damping
and process variable
damping
When
mA Output Process Variable
is set to density or temperature,
Added Damping
interacts
with
Density Damping
or
Temperature Damping
.
Example: Damping interaction
Configuration:
•
mA Output Process Variable
=
Density
•
Density Damping
= 1 second
•
Added Damping
= 2 seconds
Result: A change in density will be reflected in the mA output over a time period that is
greater than 3 seconds. The exact time period is calculated by the transmitter according to
internal algorithms which are not configurable.
Related information
6.2.4
Configure
mA Output Fault Action
and
mA Output Fault Level
Display
Not available
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
mA Output 1
>
Fault Action
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
mA Output 2
>
Fault Action
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
mA Output 1
>
mAO1 Fault Settings
>
MAO1 Fault Action
Configure
>
Manual Setup
>
Inputs/Outputs
>
mA Output 2
>
MAO2 Fault Settings
>
MAO2 Fault Action
Integrate the meter with the control system
88
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...