Example: Defining a special measurement unit for density
You want to measure density in ounces per cubic inch.
1.
Set
Density Special Unit Base
to
g/cm3
.
2.
Calculate
Density Special Unit Conversion Factor
: 1 g/cm3 = 0.578 oz/in3
3.
Set
Density Special Unit Conversion Factor
to
0.578
.
4.
Set
User-Defined Label
to
oz/in3
.
4.2.2
Configure
Density Damping
Display
Not available
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Line Density
>
Density Damping
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Density
>
Density Damping
Overview
Density Damping
controls the amount of damping that will be applied to the line density
value.
Damping is used to smooth out small, rapid fluctuations in process measurement.
Damping
Value
specifies the time period (in seconds) over which the transmitter will spread changes
in the process variable. At the end of the interval, the internal value will reflect 63% of the
change in the actual measured value.
Tip
Density damping affects all process variables that are calculated from line density.
Procedure
Set
Density Damping
to the value you want to use.
The default value is 0 seconds. The range is 0 to 60 seconds.
Interaction between
Density Damping
and
Added Damping
When the mA output is configured to report density, both
Density Damping
and
Added
Damping
are applied to the reported density value.
Density Damping
controls the rate of change in the value of the process variable in
transmitter memory.
Added Damping
controls the rate of change reported via the mA
output.
If
mA Output Process Variable
is set to
Density
, and both
Density Damping
and
Added Damping
are
set to non-zero values, density damping is applied first, and the added damping
calculation is applied to the result of the first calculation. This value is reported over the
mA output.
Configure process measurement
Configuration and Use Manual
25
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...