Sample calibration certificate: A format
Figure A-1:
DENSITY CALIBRATION CERTIFICATE
CDM100PA18MBAZ00EANZZ
2012.10031
Model Code:
Customer Tag:
Serial No:
Calibration Date:
Pressure Test:
2013-07-01 11:05:22
Calibration ID:
1.29825731
3263 psi(g)
A1 =
2.803826E-004
Time Period Temperature Compensation Coefficient:
Density Calibration Coefficients:
Case Temperature Compensation Coefficient:
Pressure Compensation Coefficient:
A2 =
-2.007
A3 =
3.083073E-005
A4=
3.21
A5=
1.207270E-001
Calibrated Range = 0.300-1.300 g/cm³ (300-1300 kg/m³)
Flow Compensation Coefficient:
A6=
3.23
Time Period, TP sec
μ
Density (g/cm³)
Density Calibration Data:
5065.2054
0.00097
5587.3151
0.79882
5857.0031
1.24079
Performance Check:
Error (g/cm³)
Density (g/cm³)
0.99822
-0.00007
Known Density Verification Data:
765.1
Verification Time Period (Air) @ 20°C =
μsec
CALIBRATED BY:
Density
Time period
Pressure
Definitions:
Temperature
g/cm³
µsec
psi(g)
°C
"A" Coefficients
Form Rev: 1
2013-08-29 13:55:05
Printed
Page 1 of 3
All equipment used for this calibration is calibrated at routine intervals against
standards that are traceable to the International System of Units (SI).
Emerson Process Management Micro Motion, Inc. 7070 Winchester Circle, Boulder, CO 80301
SAMPLE
Calibration certificate
168
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...