Overview
mA Output Fault Action
controls the behavior of the mA output if the transmitter encounters
an internal fault condition.
Note
For some faults only: If
Fault Timeout
is set to a non-zero value, the transmitter will not implement the
fault action until the timeout has elapsed.
Procedure
1.
Set
mA Output Fault Action
to the desired value.
The default setting is
Downscale
.
Restriction
If
Digital Communications Fault Action
is set to
NAN
(not a number), you cannot set
mA Output Fault
Action
to
None
. If you try to do this, the device will not accept the configuration.
2.
If you set
mA Output Fault Action
to
Upscale
or
Downscale
, set
mA Output Fault Level
as
desired.
Postrequisites
CAUTION!
If you set
mA Output Fault Action
to
None
, be sure to set
Digital Communications Fault Action
to
None
. If you
do not, the output will not report actual process data, and this may result in measurement
errors or unintended consequences for your process.
Options for
mA Output Fault Action
and
mA Output Fault Level
Options for
mA Output Fault Action
and
mA Output Fault Level
Table 6-3:
Option
mA output behavior
mA Output Fault Level
Upscale
Goes to the configured fault level
Default: 21.5 mA
Range: 21.0 to 21.5 mA
Downscale
(default)
Goes to the configured fault level
Default: 3.2 mA
Range: 3.2 to 3.6 mA
Internal Zero
Goes to the mA output level associated
with a process variable value of 0 (zero),
as determined by
Lower Range Value
and
Upper Range Value
settings
Not applicable
None
Tracks data for the assigned process vari-
able; no fault action
Not applicable
Integrate the meter with the control system
Configuration and Use Manual
89
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...