2.
Test the discrete output(s).
a. Choose
OFFLINE MAINT
>
SIM
>
DO SIM
, and select
SET ON
.
Dots traverse the display while the output is fixed.
b. Verify the signal at the receiving device.
c. At the transmitter, activate
Select
.
d. Scroll to and select
SET OFF
.
e. Verify the signal at the receiving device.
f. At the transmitter, activate
Select
.
3.
Test the TPS output.
a. Attach a frequency counter, oscilloscope, digital multimeter (DMM), or digital
voltmeter (DVM) to the TPS output loop.
b. Compare the reading to the Sensor Time Period process variable at your meter.
Postrequisites
•
If the mA output readings are within 200 microamps of each other, you can correct
this discrepancy by trimming the output.
•
If the discrepancy between the mA output readings is greater than 200 microamps,
or if at any step the reading was faulty, verify the wiring between the transmitter
and the remote device, and try again.
•
If the discrete output readings are reversed, check the setting of
Discrete Output
Polarity
.
•
If the TPS output readings do not match, contact Micro Motion customer service.
10.4.2
Perform loop tests using ProLink III
A loop test is a way to verify that the transmitter and the remote device are
communicating properly. A loop test also helps you know whether you need to trim mA
outputs.
Prerequisites
Before performing a loop test, configure the channels for the transmitter inputs and
outputs that will be used in your application.
Follow appropriate procedures to ensure that loop testing will not interfere with existing
measurement and control loops.
Procedure
1.
Test the mA output(s).
a. Choose
Device Tools
>
Diagnostics
>
Testing
>
mA Output 1 Test
or
Device Tools
>
Diagnostics
>
Testing
>
mA Output 2 Test
.
b. Enter
4
in
Fix to:
.
Troubleshooting
Configuration and Use Manual
139
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...