c. Click
Fix mA
.
d. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
e. Click
UnFix mA
.
f. Enter
20
in
Fix to:
.
g. Click
Fix mA
.
h. Read the mA current at the receiving device and compare it to the transmitter
output.
The readings do not need to match exactly. If the values are slightly different,
you can correct the discrepancy by trimming the output.
i. Click
UnFix mA
.
2.
Test the discrete output(s).
a. Choose
Device Tools
>
Diagnostics
>
Testing
>
Discrete Output Test
.
b. Set
Fix To:
to
ON
.
c. Verify the signal at the receiving device.
d. Set
Fix To:
to
OFF
.
e. Verify the signal at the receiving device.
f. Click
UnFix
.
3.
Test the TPS output.
a. Attach a frequency counter, oscilloscope, digital multimeter (DMM), or digital
voltmeter (DVM) to the TPS output loop.
b. Compare the reading to the Sensor Time Period process variable at your meter.
Postrequisites
•
If the mA output readings are within 200 microamps of each other, you can correct
this discrepancy by trimming the output.
•
If the discrepancy between the mA output readings is greater than 200 microamps,
or if at any step the reading was faulty, verify the wiring between the transmitter
and the remote device, and try again.
•
If the discrete output readings are reversed, check the setting of
Discrete Output
Polarity
.
•
If the TPS output readings do not match, contact Micro Motion customer service.
Troubleshooting
140
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...