Alert num-
ber
Alert title
Possible cause
Recommended actions
N/A
Zero Calibration in
Progress
A zero calibration is in progress.
•
No action required.
10.7
Density measurement problems
Density measurement problems and recommended actions
Table 10-2:
Problem
Possible causes
Recommended actions
Erratic density reading
•
Normal process noise
•
Two-phase flow
•
Flow rate too high
•
Contaminants or suspended solids in the
process fluid
•
Vibration in the pipeline
•
Erosion or corrosion
•
Check your process conditions.
•
Increase the density damping value.
•
Reduce the flow rate.
•
Check for two-phase flow. See
.
•
Ensure that line pressure or sample pres-
sure meets installation requirements.
•
Increase back pressure to minimize bubble
formation.
•
Minimize vibration in the pipeline.
•
Install a flow control method (bypass, flow
chamber, expander, etc.).
•
Perform Known Density Verification.
Inaccurate density
reading
•
Inaccurate temperature measurement
•
Incorrect calibration factors
•
Contaminants or suspended solids in the
process fluid
•
Verify the temperature reading from the
RTD (on-board temperature sensor).
•
Verify the temperature reading from the
external temperature device, if applicable.
•
Verify or adjust the density offset and/or
the density meter factor.
•
Increase the flow rate.
•
Install a thermal insulation jacket.
•
Verify the calibration factors.
•
Perform Known Density Verification.
Density reading too
high
•
Plugged or coated flow tube
•
Inaccurate temperature measurement
•
Incorrect calibration factors
•
Erosion or corrosion on the sensor tubes
•
Purge the flow tubes.
•
Check for coating on the flow tubes.
•
Verify the temperature reading from the
RTD (on-board temperature sensor).
•
Verify the temperature reading from the
external temperature device, if applicable.
Troubleshooting
Configuration and Use Manual
149
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...