4
Configure process measurement
Topics covered in this chapter:
•
Verify the calibration factors
•
Configure line density measurement
•
Configure temperature measurement
•
•
Configure velocity measurement
•
Set up the API referral application
•
Set up concentration measurement
4.1
Verify the calibration factors
Display
Not available
ProLink III
Device Tools
>
Calibration Data
Field Communicator
Configure
>
Manual Setup
>
Calibration Factors
Overview
The calibration factors are used to adjust measurement for the unique traits of the sensor.
Your device was calibrated at the factory. However, you should verify that the calibration
factors that are configured in your device match the factory values.
Prerequisites
You will need the factory values for the calibration factors. These are provided in two
locations:
•
The calibration certificate shipped with your meter
•
The label inside the transmitter's end-cap
Important
If the transmitter is not the original component, do not use the values from the transmitter label.
Procedure
1.
View the calibration factors that are stored in the device.
2.
Compare them to the factory values.
•
If the values match, no action is required.
•
If the values do not match, contact Micro Motion customer service.
Configure process measurement
Configuration and Use Manual
21
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...