Option
Description
mA output
Channel B will operate as the secondary mA output.
Discrete output
Channel B will operate as a discrete output.
6.2
Configure the mA output
The mA output is used to report the configured process variable. The mA output
parameters control how the process variable is reported.
The CDM mA device has two mA outputs: Channel A and Channel B. Both outputs are fully
configurable.
The CDM DO device has one mA output: Channel A. The output is fully configurable.
The CDM TPS device has one mA output: Channel A. The output is partially configurable.
Important
Whenever you change an mA output parameter, verify all other mA output parameters before
returning the meter to service. In some situations, the transmitter automatically loads a set of stored
values, and these values may not be appropriate for your application.
•
(Section 6.2.1)
•
(Section 6.2.2)
•
(Section 6.2.3)
•
(Section 6.2.4)
6.2.1
Configure
mA Output Process Variable
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CONFIG MAO 1
>
AO 1 SRC
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CH B
>
MAO 2
>
CONFIG MAO 2
>
AO 2 SRC
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
mA Output 1
>
Source
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
mA Output 2
>
Source
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
mA Output 1
>
Primary Variable
Configure
>
Manual Setup
>
Inputs/Outputs
>
mA Output 2
>
Secondary Variable
Overview
Use
mA Output Process Variable
to select the variable that is reported over the mA output.
Procedure
Set
mA Output Process Variable
as desired.
Default settings are shown in the following table.
Integrate the meter with the control system
84
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...